Scientech Corporation  
  繁中
   
 

 
 
   
Product
   
Chemistry and Biotechnology
arrow2 Acromass
arrow2 Berghof
arrow2 CAMAG
arrow2 Echo
arrow2 Eltra
arrow2 Fritsch
arrow2 KLA
arrow2 KLab
arrow2 Labex
arrow2 M Napra
arrow2 MOCON
arrow2 Neutronix Quintel
arrow2 Otsuka
arrow2 Pickering
arrow2 Rubix
arrow2 Shodex
arrow2 Sinyang
arrow2 Teledyne
Leeman Labs
arrow2 Teledyne
Tekmar
arrow2 ThermoFisher
arrow2 TMC
arrow2 ULVAC
arrow2 WGE
arrow2 Zeiss
Broadcast Equipment
Semiconductor III-V Group
LCD/ LED/ SOLAR/ DATA STORAGE
MEMS
PCB Industry
Packaging & Testing
Used Tool
Scientech Tool
3D Printer
  Chemistry and Biotechnology - Fluid Imaging Technologies - FlowCam ALH-Automated Liquid Handling System



FlowCam ALH-Automated Liquid Handling System
The FlowCam® Automated Liquid Handling system from Fluid Imaging Technologies integrates with the FlowCam® 8100 for uninterrupted processing of samples.
The system allows up to 96 samples to be queued for unattended operation, improved repeatability, and increased productivity。
Main Features:
icon 1. Pre-analysis sample preparation can be programmed using easy to use graphical and menu driven operations
2. Heating & cooling features allow sample conditioning and control of sample degradation and evaporation during the processing.
3. System is self-contained to protect against outside contamination and accidental disturbance of samples.
4. Vertical architecture saves valuable bench space while a monitor provides access to current system status and review of workflows for both instruments.Uses standard SBS format labware that can deployed on any of the general purpose locations.
Standard Certification:
Application:
Specification:
Introduction video:
FlowCam Automated Liquid Handling for Pharmaceutical and Biopharma
https://www.youtube.com/watch?v=5FFblBeJV4w

FlowCam ALH Specifications

Operating principle

Single robotic XYZ arm, hydraulic fill on all fluidics

Labware format

SBS

Number of positions for labware

7

Maximum number of wells per deck position

96

Maximum volume dispensed per operation

1000μl

Minimum volume dispensed during dilution or reagent

5μl

Minimum volume dispensed into FlowCam for analysis

200μl

CV of aspiration/dispensing accuracy

±0.5%

Pipette format

Disposable, Tecan® style tips

Environmental (particulate) control

Positive pressure HEPA filtration

Sterilization

UV, programmable exposure time

Mechanical orbital method

100 -2000 RPM clockwise

Sample conditioning

Heating/cooling 2-90°C

Mixing

Aspiration/dispensing method

Multiple assays/workflows supported

Yes

Operation pause and resume

Yes

CFR21 Compliant

Yes

Workspace illumination

User controlled white LED

Size

Height: 49in. (124.5 cm), Width: 30 in. (76.2 cm), Depth:25 in (63.5 cm)

Minimum bench space required

Height: 55 in. ( 140 cm), Width: 60 in. (153 cm), Depth:30in (77 cm)

Weight

190 lb. (86.7 kg)

Voltage

110-220

Power

700 watts, including FlowCam

Operating Temperature

Operating: 5-40° C, Humidity 20-80% RH

Storage Temperature

Storage: -20-60° C, Humidity 20-95% RH

 

 
 
 
 
 
Copyright © 2009 SCIENTECH. All rights reserved. Site map